EasyManua.ls Logo

Siemens SONOLINE G50 - Area Measurements on a Planar Quadrant

Siemens SONOLINE G50
229 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6 T e chnical Description
[ 1 ] I N S T R U C T I O N S F O R U S E 6 - 1 9
Area Measurements on a Planar Quadrant
Note: Area measurements are only possible on a planar quadrant (slices). You cannot
perform an area measurement on the volume quadrant.
The measurement accuracy tables are based on the worst-case linear
measurements for a planar quadrant.
Transducer: C6-3 3D, C6F3
NTSC Systems, Standard Sweep Speed
Measured Area
Depth
20 x 50 mm 20 x 60 mm 50 x 60 mm
80 mm within ± 11.1 % within ± 11.1 % within ± 10.3 %
160 mm within ± 16.1 % within ± 16.1 % Within ± 10.3 %
240 mm within ± 28.3 % within ± 28.3 % Within ± 15.4 %
Transducer: C6-3 3D, C6F3
NTSC Systems, Medium Sweep Speed
Measured Length
Depth
20 mm 50 mm 60 mm
80 mm within ± 16.8 % within ± 16.8 % within ± 10.3 %
160 mm within ± 28.1 % within ± 28.1 % within ± 13.9 %
240 mm within ± 57.8 % within ± 57.8 % within ± 30.8 %
Transducer: C6-3 3D, C6F3
PAL Systems, Standard Sweep Speed
Measured Length
Depth
20 mm 50 mm 60 mm
80 mm within ± 11.1 % within ± 11.1 % within ± 10.3 %
160 mm within ± 16.1 % within ± 16.1 % within ± 10.3 %
240 mm within ± 34.0 % within ± 34.0 % within ± 18.4 %
Transducer: C6-3 3D, C6F3
PAL Systems, Medium Sweep Speed
Measured Length
Depth
20 mm 50 mm 60 mm
80 mm within ± 16.8 % within ± 16.8 % within ± 10.3 %
160 mm within ± 28.1 % within ± 28.1 % within ± 13.9 %
240 mm within ± 70.6 % within ± 70.6 % within ± 37.4 %