6 T e chnical Description
[ 1 ] I N S T R U C T I O N S F O R U S E 6 - 1 9
Area Measurements on a Planar Quadrant
Note: Area measurements are only possible on a planar quadrant (slices). You cannot
perform an area measurement on the volume quadrant.
The measurement accuracy tables are based on the worst-case linear
measurements for a planar quadrant.
Transducer: C6-3 3D, C6F3
NTSC Systems, Standard Sweep Speed
Measured Area
Depth
20 x 50 mm 20 x 60 mm 50 x 60 mm
80 mm within ± 11.1 % within ± 11.1 % within ± 10.3 %
160 mm within ± 16.1 % within ± 16.1 % Within ± 10.3 %
240 mm within ± 28.3 % within ± 28.3 % Within ± 15.4 %
Transducer: C6-3 3D, C6F3
NTSC Systems, Medium Sweep Speed
Measured Length
Depth
20 mm 50 mm 60 mm
80 mm within ± 16.8 % within ± 16.8 % within ± 10.3 %
160 mm within ± 28.1 % within ± 28.1 % within ± 13.9 %
240 mm within ± 57.8 % within ± 57.8 % within ± 30.8 %
Transducer: C6-3 3D, C6F3
PAL Systems, Standard Sweep Speed
Measured Length
Depth
20 mm 50 mm 60 mm
80 mm within ± 11.1 % within ± 11.1 % within ± 10.3 %
160 mm within ± 16.1 % within ± 16.1 % within ± 10.3 %
240 mm within ± 34.0 % within ± 34.0 % within ± 18.4 %
Transducer: C6-3 3D, C6F3
PAL Systems, Medium Sweep Speed
Measured Length
Depth
20 mm 50 mm 60 mm
80 mm within ± 16.8 % within ± 16.8 % within ± 10.3 %
160 mm within ± 28.1 % within ± 28.1 % within ± 13.9 %
240 mm within ± 70.6 % within ± 70.6 % within ± 37.4 %