Transducer Surface Temperature Limits
The following table provides the maximum surface temperature of the
transducers compatible with the system.
Maximum surface temperatures are in accordance with IEC 60601-2-37.
Maximum Temperature
Transducer TMM Still Air
C5-2
d 41.0q C d 41.7q C
C6-2
d 41.0q C d 41.4q C
C6-3 3D/C6F3
d 41.0q C d 41.0q C
C8-5
d 41.0q C d 41.0q C
5.0C50+
d 41.0q C d 41.3q C
MPT7-4
d 41.0q C d 41.0q C
BE9-4
d 41.4q C d 41.9q C
EV9-4
d 41.0q C d 41.0q C
EC9-4
d 41.0q C d 41.0q C
Endo-V II
d 41.0q C d 41.0q C
Endo-V 3D/EV8F5
d 41.0q C d 41.0q C
5.0L45
d 41.0q C d 41.0q C
7.5L70
d 41.0q C d 41.0q C
L10-5
d 41.0q C d 41.0q C
VF13-5
d 41.1q C d 41.0q C
VF13-5SP
d 41.8q C d 42.9q C
P4-2
d 41.0q C d 41.0q C
5.0P10
d 41.1q C d 41.6q C
CW2
d 41.0q C d 41.0q C
CW5
d 41.0q C d 41.0q C
TMM = Tissue Mimicking Material