SDS1000CML/CNL/DL Service Manual 25
15. Set the amplitude of the Sine Wave Generator to 750mVpp, so that
the Vpp measurement of the oscilloscope is approximately 750mv.
16. Press Trigger LEVEL button to move the waveform to the screen
center.
17. Adjust the Trigger LEVEL within the waveform range to check if
there is stable trigger.
18. Disconnect the test connection.
19. Check CH1 in the same way as step 1 to 18.