SIGLENT
26 SDS1000CML/CNL/DL Service Manual
Adjusting Procedures
This chapter explains how to adjust the SDS1000CML/CNL/DL series
oscilloscope with FLUKE 9500B for optimum operating performance. Only
qualified personnel should perform this procedure.
Warming up
Before performing the adjustment procedures, you must let the oscilloscope
and other test equipments warm up for at least 30 minutes in an ambient
temperature between 20 °C and 30 °C. Adjustments performed prior to
warm-up or outside this temperature range may result in poor performance.
Self calibration
The Self-Cal performs an internal routine which uses internally generated
signals to optimize circuits that affect channel sensitivity, offset, and trigger
parameters, to optimize the signal path in the oscilloscope. Let the
oscilloscope warm up before performing this procedure.
Required Equipments
PC: Inter Pentium processor, Windows XP operating system, USB interface,
Microsoft Office Excel 2003, NI-GPIB Card and corresponding NI-488.2
driver
Calibrating Software: EasyTest, Adjustment script
Test Instrument: FLUKE 9500B with active head
An USB cable (Type AB)
A GPIB cable
Two 10-inch BNC cables
A BNC T connector
Software Installation
1. Install the NI-488.2 driver, select custom installation to install NI-VISA.
2. Install the EasyTest software.
3. Decompress production. zip file to E disk directory.
Adjusting steps
1. Connect the FLUKE 9500B with PC using a GPIB cable.
2. Connect the oscilloscope with PC using an USB cable.