9 Technical data
112 SOREDEX
206718 rev 9
INSTALLATIONS REQUIREMENTS & ENVIRONMENT
CONTROL:
In order to minimize interference risks, the following
requirements shall apply.
Cables shielding & grounding
All interconnect cables to peripheral devices must be
shielded and properly grounded. Use of cables not
properly shielded and grounded may result in the
equipment causing radio frequency interference.
Electrostatic discharges environment & recommendations
In order to reduce electrostatic discharge interference, a
charge dissipative floor should be installed to prevent
charge accumulation.
■ The dissipative floor material must be connected to
the system reference ground, if applicable.
■ Relative humidity must be maintained above 30
percent.
Stacked components & equipment
The CRANEX 3D should not be used adjacent to or
stacked with other equipment; if adjacent or stacked use is
necessary, the CRANEX 3D should be observed to verify
normal operation in the configuration in which it will be
used.
Interference may occur in the vicinity of equipment
marked with the following symbol:
No portable or mobile RF communications equipment may
be used closer to any part of the CRANEX 3D including
cables, than the recommended separation distance
calculated from the equation appropriate to the frequency
of the transmitter.
See Table 4.