EasyManua.ls Logo

Soredex CRANEX 3D - 3 D Exposure Settings and AES

Default Icon
126 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3 Imaging programs
206718 rev 9
SOREDEX 27
3.7 Exposure settings for 3D imaging
Exposure settings for 3D imaging (default values)
NOTICE!
Voltage setting is always 90kV with the unit 3D modality.
FOV Resolution Exposure
time
Scanning
time
Voxel
size
(µm)
Default
mA
mAs DAP
(mGycm2)
61 x 41 mm
Minidose 1,2 s 11,0 s 290 3.2 3,8 30
Std Res 2,3 s 15,6 s 200 10 23 184
High Res 6,1 s 11,5 s 133 8 48,8 384
Endo Res 6,1 s 11,5 s 85 10 61 480
61 x 78 mm
(optional)
Minidose 2,4 s 21,1 s 330 3.2 7.7 61
Std Res 4,9 s 21,1 s 300 10 49 383
High Res 12,6 s 21,0 s 200 6.3 79,4 626
Available mA ranges for each field of view sizes and resolution settings
FOV Resolution 3.2
mA
3.6
mA
4
mA
4.5
mA
5
mA
5.6
mA
6.3
mA
7.1
mA
8
mA
9
mA
10
mA
11
mA
12.5
mA
14
mA
61 x 41 mm
Minidosexxxxxxxx
Std Res xxxxxxx
High Res xxxxxxxxxxx
Endo Res xxxxxxxxxxx
61 x 78 mm
(optional)
Minidose x x x x x
Std Res xxxxxxx
High Res xxxxxxxxx
Exposure settings for scout imaging (default values)
FOV
(h x w)
Resolution kV mA Scan-
ning
time
61 x 41 mm Scout 90 12.5 0,02 s
61 x 78 mm
(optional)
Scout 90 12.5 0,04 s