8 Calibration and adjustment
82 SOREDEX
206718 rev 9
8.2 Preparing for calibration
1. Close the head support and lock it in its upmost position.
2. Switch the PC and unit on.
3. PC: Open the dental imaging software and then open a
patient (card) and give it an identifiable name, for exam-
ple: calibration (refer to the user’s manual supplied with
the dental imaging software for more information).
4. PC: Click the image acquisition button to activate image
capture.
5. Touch the settings button on the touch screen display.
6. Select the Quality assurance button.
The calibration display appears.