3 Imaging programs
28 SOREDEX
206718 rev 9
3.8 Automatic Exposure setting (AES)
AES automatically recommends the exposure values
according to the size of the patient’s head in panoramic
and cephalometric imaging. The AES is activated with
active A button and by pushing Return button once. The
AES suggest exposure values based on patient size. The
user can change the value before exposure if needed by
pushing A button again and changing the values.
NOTICE! Values can be re-scaled by SOREDEX
authorized service personnel.
3.8.1 AES values for Panoramic imaging
Adult, child panoramic, Bitewing, Sinus
Head width (min) kV mA
19,3 cm (Max) 70 12.5
15,9 cm 70 10
14,2 cm 70 8
13,35 cm 66 8
12,925 cm 66 6,3
12,5 cm (Min) 66 5
AES values for TMJ Lat/PA
Head width (min) kV mA
19,3 cm (Max) 73 16
15,9 cm 73 12.5
14,2 cm 73 10
13,35 cm 73 8
12,925 cm 73 6,3
12,5 cm (Min) 73 5