9 Technical data
108 SOREDEX
206718 rev 9
9.5 Electromagnetic Compatibility (EMC)
tables
NOTICE! Medical electrical equipment needs special
precautions regarding EMC and needs to be installed
according to EMC information.
Table 1.1 Electromagnetic emissions IEC 60601-1-2 Ed3
CRANEX 3D is suitable for use in the specified electromagnetic environment. The purchas-
er or user of CRANEX 3D should assure that it is used in an electromagnetic environment
as described below:
Emissions Test Compliance Electromagnetic Environment
Radio-Frequency
Emissions
CISPR11
Group 1 CRANEX 3D uses RF energy only for its internal
function. Therefore, the RF emission is very low and not
likely to cause any interference in nearby electronic
equipment.
Radio-Frequency
Emissions
CISPR11
Class B CRANEX 3D is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.
Harmonic
emissions
IEC 61000-3-2
IEC 61000-3-2
Class A
CRANEX 3D is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.
Voltage
fluctuations/
flicker emissions
IEC 61000-3-3
Complies CRANEX 3D is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.