9 Technical data
206718 rev 9
SOREDEX 109
Table 1.2 Electromagnetic immunity IEC 60601-1-2 Ed3
CRANEX 3D is suitable for use in the specified electromagnetic environment. The purchas-
er or user of CRANEX 3D should assure that it is used in an electromagnetic environment
as described below:
Immunity Test IEC 60601-1-2
Test Level
Compliance
Level
Electromagnetic Environment
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 2, 4, 6 kV for
contact discharge
± 2, 4, 8 kV for air
discharge
± 2, 4, 6 kV for
contact discharge
± 2, 4, 8 kV for air
discharge
Floors are wood, concrete, or
ceramic tile, or floors are covered
with synthetic material and the
relative humidity is at least 30
percent.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for
input/output
lines
± 2 kV for power
supply lines
± 1 kV for
input/output
lines
Mains power quality is that of a
typical commercial and/or hospital
environment
Surge
IEC 61000-4-5
± 1 kV differential
mode
± 2 kV common
mode
± 1 kV differential
mode
± 2 kV common
mode
Mains power quality is that of a
typical commercial and/or hospital
environment.
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines
IEC 61000-4-11
< 5 % UT
(> 95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
< 5 % UT
(> 95 % dip in UT)
< 5 % UT
(> 95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
< 5 % UT
(> 95 % dip in UT)
Mains power quality is that of a
typical commercial and/or hospital
environment. If the user of CRANEX
3D requires continued operation
during power mains interruptions, it
is recommended that CRANEX 3D
be powered from an uninterruptible
power supply or a battery.
Power
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic fields are
at levels characteristic of a typical
location in a typical commercial and/
or hospital environment.
NOTE: U
T
is the a.c. mains voltage prior to application of the test level.