3 Imaging programs
206718 rev 9
SOREDEX 29
3.8.2 AES values for Cephalometric imaging
Lat/PA 100 V
Head width (min) kV mA s
17,3 cm (Max) 90 8 20
14,45 cm 90 8 20
12,65 cm 90 8 16
11,45 cm 90 8 12,5
10,55 cm 90 8 10
6,5 cm (Min) 85 8 10
Ceph Lat/PA 120 V
Head width (min) kV mA s
17,3 cm (Max) 90 10 20
14,45 cm 90 10 20
12,65 cm 90 10 16
11,45 cm 90 10 12,5
10,55 cm 90 10 10
6,5 cm (Min) 85 8 10
Ceph Lat/PA 240 V
Head width (min) kV mA s
17,3 cm (Max) 90 12,6 16
14,45 cm 85 12,6 16
12,65 cm 85 12,6 12,5
11,45 cm 85 12,6 10
10,55 cm 85 10 10
6,5 cm (Min) 85 8 10