8 Calibration and adjustment
88 SOREDEX
206718 rev 9
8.4.2 3D pixel calibration
1. Remove the 3D calibration phantom.
2. Select the program.
3. Press Return.
4. Take an exposure. The result image informs when the
calibration is passed.