7 Maintenance
80 SOREDEX
206718 rev 9
Steam sterilization
Recommended parameters for sterilizable parts are:
Gravity-displacement steam sterilization
"Flash" sterilization:
Temperature: 270 F (132°C)
Exposure time: 3 minutes
Prevacuum steam sterilization
"Flash" sterilization:
Temperature: 270 F (132°C)
Exposure time: 3 minutes
Steam-flush pressure-pulse steam sterilization
Temperature: 270 F to 275 F (132°C to 135°C)
Exposure time: 3 to 4 minutes