Chapter 9 : Recipes
9-10 Lasair® Pro Aerosol Particle Counter Operations Manual
Recipe Importing and Exporting
Lasair Pro recipes can be imported and exported to other Lasair Pro counters. See Clone
Export and Import on page 12-8. Be aware of the following:
Moving a recipe between instruments with dierent flowrates may result in
unexpected sample times or sample volumes. The Lasair Pro can sample by time
or by volume, and this information is recorded in a sample recipe.
For example, if a one minute sample recipe is exported from a 310 model and
imported to a 5100, then the sample volume on the 5100 will be 100 liters rather
than 1 , in order to keep the time the same.
Moving a recipe between instruments with dierent channel sizes may result in
missing alarms.
For example, a recipe created on a 310 model with an alarm of 1000 counts @ 0.3
m will be unable to export that alarm setting properly to a 5100, where there is no
0.3 m channel.
Moving a statistics recipe between instruments with dierent channel sizes may
result in some adjustments to the statistics if the channel size is not supported in
the new instrument.
Recipe Details
You can view the details of saved recipes on the SETUP tab of the RECIPES screen by
pressing . A typical recipe includes the following details:
Sample type (volume or time)
Sample size (volume or time)
Initial purge delay (in seconds)
Start by time of day (On/O)
Inter-sample delay (in seconds)
Number of samples
Continuous sampling (On/O)
Group repeat mode (On/O)

Table of Contents