5-15Lasair® Pro Aerosol Particle Counter Operations Manual
Chapter 5 : Sampling Options
Wall/Hand/Tripod Adapters
A wall/hand/tripod adapter, specific to the particle counter model, is included and
shipped with each Lasair Pro particle counter. Additional adapters are available for
purchase, if needed:
• Wall/hand/tripod adapter (1 CFM) – P/N 1000012955
• Wall/hand/tripod adapter (50 LPM) – P/N 1000012956
• Wall/hand adapter (100 LPM) – P/N 1000012957
Zero Count Filters
• Zero count filter (1 CFM) – P/N 90104050
• Zero count filter (50 LPM) – P/N 90104052
• Zero count filter (100 LPM) – P/N 90104055
Thermal Paper
• Standard thermal paper – P/N 1000024998
• Cleanroom thermal paper – P/N 1000024977
For users who desire to use the Isokinetic Sampling Probe (ISP) remotely with a handheld
accessory, there is the option of using a handheld isokinetic probe attachment (HHIPA).
The HHIPA is ideally designed for filter scanner applications where the standard ISP
is not suitable. The HHIPA meets the requirements of ISO 14644-3, Section B.6.2.4,
“Determination of Probe Size”, and can be used for filter scanning applications.
HHIPA Accessory
Flowrate
The HHIPA probe is available in three dierent flowrate configurations, depending on the
flowrate of your Lasair Pro particle counter. Those flowrates are:
• 28.3 LPM (1.0 CFM)
• 50 LPM