11-1Lasair® Pro Aerosol Particle Counter Operations Manual
Overview
The Lasair Pro particle counter can be configured for certification statistics when in
Statistics Mode. A choice of certification standards includes ISO 14644-1:2015, FS-209E, EU
GMP, and AVERAGE.
Selecting the certification standard and class automatically sets the particle count limits
for the sampling location. Aer configuring parameters such as the size of the room to
be certified and its state (dynamic or static), enabling Statistics Mode opens a prompt
window for the user to select their locations to sample. The user is automatically taken
to the HOME screen to start the sampling process, where upon successful completion, a
report is generated.
Figure 11-1. Example Cleanroom Certification Report