Chapter 9 : Recipes
9-12 Lasair® Pro Aerosol Particle Counter Operations Manual
Statistics Recipe Details
Statistics recipes are found in the same list as normal recipes. Statistics recipes are saved
on the SETUP tab of the STATISTICS screen. See SETUP Tab on page 11-9. The following
is the list of all parameters saved for a statistics recipe:
• Channels configuration (List of channels included in statistics plan)
• Instrument type (Lasair Pro model)
• Start type (Standard or )
• Statistics standard (ISO-14644-1 2015, EU GMP, FS-209E, or AVERAGE)
• Class (not applicable for AVERAGE mode)
• Purge delay (seconds)
• Inter-sample delay (seconds)
• Auto-Track mode (On/O)
• Confirm sample mode (On/O)
• Units (
3
/m
3
)
• Particle channel
• Flow (Uni-directional/Multi-directional)
• Room state (Static or Dynamic)
• Room area
• Room area units
• Number of locations to sample
• Volume per sample
• Number of samples per location
Statistics recipe data not saved:
• Statistics User Name
• Sampling Setup Screen Parameters – The number of samples is automatically set
when statistics are enabled, except for average mode. In average mode, the current
sample count is used.
• Alarm settings – Default alarms are disabled during statistics.