1-8 Lasair® Pro Aerosol Particle Counter Operations Manual
Using NFC
NFC cards can store one user’s credentials or one recipe configuration for ease of use. To
enable NFC:
1. Press , then (COMM).
2. Go to the NFC tab and press the Enable Security, Enable Locations/Recipes,
and Enable Write NFC indicators. A gray dot indicates the setting has not been
enabled.
3. When on the USERS tab of the SECURITY screen, and the ASSIGNMENTS tab of the
RECIPES screen, the NFC icon
and options will be activated.
Data Export and Import
The Lasair Pro system and data can be cloned and copied to another Lasair Pro
instrument. You can export a clone file or import an existing one. Settings in the clone file
include:
• All SYSTEM screen parameters
• All SAMPLE screen parameters
• All STATISTICS screen parameters
• All SECURITY screen parameters
(including all user data)
• TCP/IP communication parameters
• All COMM screen parameters
• IP addresses (selectable with Comm
Settings)
• Area, Room, Locations and
assignments
• Recipes
• All IN/OUT screen parameters
• All DISPLAY screen parameters
• All DATA screen parameters
• Clone file encryption
1. Press
(SYSTEM).
2. Go to the SETUP tab.