11-5Lasair® Pro Aerosol Particle Counter Operations Manual
Chapter 11 : Statistics
Figure 11-4 shows the following adjustments to the HOME screen:
• A. Average and cumulative normalized counts: Average values refer to the mean
particle count data for all samples across all locations, directly comparable to
cumulative normalized counts (counts of a certain particle size and larger over the
configured sample volume) for all locations.
• B. Average, maximum, and 95% UCL statistics: Average values refer to the mean
particle count data for all samples across all locations, directly comparable to the
maximum particle size count of all samples, and the 95% Upper Confidence Limit,
a more conservative value indicating how many particles or less of a certain size
are found within the tested locations with 95% certainty.
NOTE: 95% UCL will only be calculated for sample numbers configured as ≥ 2 and ≤ 9.
• C. Average particle counts by size and location: Organizes data into tabular
format by location and the average count for each particle size. Swipe right and le
to view the table in its entirety.
Results
You can print or save a classification report similar in scope to the other standard options
by pressing either or on the bottom-le of display B or C of the HOME
screen (see Figure 11-4). An example is provided in Figure 11-5.