EasyManua.ls Logo

Spectris PARTICLE MEASURING SYSTEMS Lasair Pro 310 - Page 32

Spectris PARTICLE MEASURING SYSTEMS Lasair Pro 310
235 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2-3Lasair® Pro Aerosol Particle Counter Operations Manual
Chapter 2 : Specifications
MODEL
310 350 5100
Enclosure
Cleaning
Materials
Alcohol, 70% or 99.5%
Hydrogen peroxide 6%
Phenol or other phenolics
Quaternary ammonium salts
Hypochlorites/sodium hypochlorite 0.5% detergent
Peracetic acid and hydrogen peroxide combination
Sample Tubing
ID
3/8-inch
8 m max. length
1/2-inch
8 m max. length
3/4-inch
8 m max. length
Sample Output
Filtering
Internally filtered to > 99.97% at 0.3 m
External AC
to DC Power
Supply
Input: 100 – 240 V, 50/60 Hz, 2.5 A
Output: 19 VDC, 7.89 A
AC input voltage fluctuation shall not exceed + 10%.
(Requires a grounded electrical supply for safety.)
Optional Battery Rechargeable Lithium Ion battery pack NH2054QE34
Estimated continuous operation is:
3 hr (single) and 6 hr (dual) for 1 CFM unit
2.5 hr (single) for 50 LPM and 3 hr for 100 LPM (dual)
Dimensions
(H x W x D)
8.5 x 10.1 x 9.6 in (21.6 x 25.6 x 24.4 cm)
Weight 11.9 lb (5.4 kg) without battery
14.2 lb (6.4 kg) with two optional batteries
Operating
Environment
Temperature: 0 – 30 °C
Humidity: 5 – 95% RH non-condensing
Table 2-1. Specifications
a. Less than 10% coincidence loss at maximum recommended concentration.
b. For pharmaceutical applications, tubing length should equal 2m maximum.

Table of Contents

Related product manuals