Circuit Descriptions
1502C MTDR Service Manual
519
50 ns
START PULSE
VAR. DELAY RAMP
TO SET DELAY ZERO
COMPARATOR LEVEL SET
SO SAMPLE TAKEN AT 10%
POINT ON OUTPUT PULSE
PULSE OUT
10% LEVEL
COMPARATOR
OUTPUT
DELAY ZERO SET
50 ns RAMP SET
FIXED CIRCUIT
DELAYS
50 ns RAMP START
0V = 0 DELAY
50 ns DELAY
COMPARATOR LEVEL
SET TO SAMPLE
PULSE AT 10% POINT
ON OUTPUT PULSE
0V COMPARATOR LEVEL = 0 DELAY
4V = 50ns DELAY
20 MHz CLK
TP2031
TRIG TO
TRIG TO
SAMPLE TRIG
PULSE TRIG
DIGITAL DELAY
PULSE GEN.
TP9011
SAMPLER
TP7010
TP2030
Figure 511: Calibration of Delay Zero and 50-ns Analog Delay