EasyManua.ls Logo

Tektronix MSO3000 Series - Page 168

Tektronix MSO3000 Series
196 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Application Exa
mples
Reference
Levels
5. Push Set Levels in to select units.
Set Levels
in
%|
Units
6. Push High Ref and turn multipurpose knob
a to enter 2.00 V. If needed, push Fine to
change the se
nsitivity o f the multipurpose
knob.
High Ref
(a) 2.00 V
Mid Ref
7. Push Low Re
f and turn multipurpose knob
a to enter 800 mV. If needed, push Fine to
change the sensitivity of the multipurpose
knob.
Low Ref
(a) 800 mV
Rise time is typically measured between the 10% and 90% amplitude levels of a signal. These are the default reference
levels th
at the oscilloscope uses for rise time measurements. However, in this example, you need to measure the time that
the signal takes to pass between the 0.8 V and 2.0 V levels.
You can customize the rise time measurement to m easure the signal transition time between any two reference levels. You
can set each of these reference levels to a specic percent of the signal amplitude or to a specic level in vertical units
(such as
volts or amperes).
Measuring specic events. Next you want to see the pulses in the incoming digital signal, but the pulse widths vary so
it is hard to establish a stable trigger. To look at a snapshot of the digital signal, do this step:
1. Push Single to capture a single acquisition.
This assumes that the oscilloscope w ill
trigger with the current settings.
Now you want to measure the width of each
displayed pulse. You can use measurement
gating to select a specic pulse to measure.
To measure the second pulse:
2. Push Measure.
146 MSO3000 and DP O3000 Series Oscilloscopes User Manual

Table of Contents

Other manuals for Tektronix MSO3000 Series

Related product manuals