Appendix: Warra
nted Speciļ¬cations
Appendix: War
ranted Speciļ¬cations
Analog bandwi
dth,
50 ā¦
The limits stated below are for ambient temperature of ā¤30 °C and the bandwidth selection set to FULL.
Reduce the upper bandwidth frequency by 1% for each °C above 30 °C.
Instrument Bandwidth
10 m V /div to
1V/div
5 mV/div to
9.98 mV/div
2mV/div to
4.98mV/div
1 mV/div to
1.99 mV/div
MSO/DPO 3054
500 MHz
DC to 500 MHz DC to 400 MHz DC to 250 MHz DC to 150 MHz
DPO 3052
500 MHz
DC to 500 MH
z
DC to 400 MH
z
DC to 250 MH
z
DC to 150 MH
z
MSO/DPO 3034
500 MHz
DC to 500 MHz DC to 400 MHz DC to 250 MHz DC to 150 MHz
300 MHz
DC to 300 MHz DC to 250 MHz DC to 150 MHz
MSO/DPO
3032
500 MHz
DC to 500
MHz
DC to 400
MHz
DC to 250
MHz
DC to 150
MHz
300 MHz
DC to 300 MHz DC to 250 MHz DC to 150 MHz
MSO/DPO 3014
500 MHz
DC to 500 MHz DC to 400 MHz DC to 250 MHz DC to 150 MHz
300 MHz
DC to 3
00 MHz
DC to 2
50 MHz
DC to 1
50 MHz
100 MHz
DC to 100 MHz
MSO/DPO 3012
500
MHz
DC to 500 MHz DC to 400 MHz DC to 250 MHz DC to 150 MHz
300 MHz
DC
to 300 MHz
DC
to 250 MHz
DC
to 150 MHz
100 MHz
DC to 100 MHz
1Mā¦: ±1% in parallel with 11.5 pF ±2 pF
I
nput impedance,
DC coupled
75 ā¦:±1%VSWR⤠1.3:1fromDCto60MHz,typical
50 ā¦:±1%
For 500MHz bandwidth models: VSWR ⤠1.5:1 from DC to 500 MHz, typical
For 300MHz bandwidth models: VSWR ⤠1.5:1 from DC to 350 MHz, typical
For 100MHz bandwidth models: VSWR ⤠1.5:1 from DC to 100 MHz, typical
DC Balance 0.2 div w ith the input DC-50 ⦠coupled and 50 ⦠terminated
0.2 div with the input DC-75 ⦠coupled and 75 ⦠terminated
0.25 div at 2 mV/div with the input DC-50 ⦠coupled and 50 ⦠terminated
0.25 div at 2 mV/div with the input DC-75 ⦠coupled and 75 ⦠terminated
0.5 div at 1 mV/div with the input DC-50 ⦠coupled and 50 ⦠terminated
0.5 div at 1 mV/div with the input DC-75 ⦠coupled and 75 ⦠terminated
0.2 div with the input DC-1 M⦠coupled and 50 ⦠terminated
0.3 div at 1 mV/div with the input DC-1 M⦠coupled and 50 ⦠terminated
DC Gain accuracy ±2.5% for Coarse setting of 1 mV/div, derated at 0.100%/°C above 30 °C
±2.0% for Coarse setting of 2 mV/div, derated at 0.100%/°C above 30 °C
±1.5% for Coarse settings o. 5 m V/div and above, derated at 0.100%/°C above 30 °C
±3.0% for all Fine V/div settings, derated at 0.100%/°C above 30 °C
MSO3000 and DPO 3000 Series Oscilloscopes User Manual 165