Installation
Operating Considerations
MSO3000 and DP
O3000 Series
Oscilloscopes
Power Supply Input Voltage: 100 V to 240 V ± 10%
Power Supply Input Power Frequency:
50/60 Hz at 100 V to 240 V
400 Hz ± 10% at 115 V
Power Consumption: 120 W maximum
Weight: 4.2 kg (9.2 lbs), standalone instrument
Height, including feet but not handle:
203.2 mm (8 in)
Width, 416.6 mm (16.4 in)
Depth, 147.4 mm (5.8 in)
Clearance: 51 mm (2 in)
MSO3000 series
Temperature:
Operating: 0 °C to +50 °C (+32 °F to +122 °F)
Nonoperating: -40 °C to +71 °C (-40 °F to +160 °F)
Humidity:
Operating: 5% to 95% relative humidity (RH) at up
to +30 °C
Operating: 5% to 45% relative humidity (RH) above
+30 °C up to + 50 °C, non-condensing, and as limited
by a Maximum Wet-Bulb Temperature of +38 °C
(derates relative humidity to 45 % RH at +50 °C)
Non-operating: 5% to 95% Relative Humidity (RH)
at up to +30 °C
Non-operating: 5% to 45% Relative Humidity (RH)
above +30 °C up to +50 °C, non-condensing, and
as limited by a Maximum Wet-Bulb Temperature of
+38 °C (derates relative humidity to 27% RH at 60 C)
DPO3000 series
Altitude:
Operating: 3,000 m (9,843 ft)
Nonoperating Altitude: 12,000 m (39,370 ft)
Random Vibration:
Operating: 0.31 G
RMS
, 5 500 Hz, 10 minutes per axis, 3 axes (30 minutes total)
Non-operating: 2.46 G
RMS
, 5 500 Hz, 10 minutes per axis, 3 axes (30 m inutes total)
Pollution Degree: 2, Indoor use only
Acquisition System: 1 M
The maximum input voltage: At front-panel connector, 300 V
RMS
, Installation Category II; for measurements performed on
circuits directly connected to the low-voltage installation.
MSO3000 and DPO 3000 Series Oscilloscopes User Manual 5