EasyManuals Logo

THORLABS ITC510 User Manual

THORLABS ITC510
123 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #12 background imageLoading...
Page #12 background image
1.3 Features
ITC500 / page 4
1.3 Features
Key-operated power switch
Protection against unauthorized or accidental use.
Defined states after switch-on
When switching on the ITC5xx with the mains switch the laser and TEC current
outputs are always switched off. The ITC5xx will always wake up in a mode that
can be selected via a DIP switch at the rear.
Softstart
The switch on delay of the softstart function protects the laser diode against
undesired transients.
Limit for the laser diode- and the TEC current (hardware limit)
Manually adjustable current limits restrict the maximum allowed current. The limit
values are set with two 25-turn potentiometers at the front panel.
Interlock
The interlock input realizes several protective functions simultaneously.
Safety lock to prevent unintentional use
Cable damage monitoring
An external emergency switch may be connected
Application of external automatic protection equipment, e.g. for the temperature
of cooling water
ON/OFF LED of the laser diode
The laser can only be operated with the interlock input being duly closed.
Contact protection of the laser diode (open circuit)
If the cable connecting the laser should be interrupted even for a short time during
operation, an emergency switch turns off laser current immediately.
Contact failure display of the TEC element (open circuit)
Cable damage, bad contact or too high resistive TEC element is displayed with a
LED. The output will remain switched on since also a wrong TEC element can still
provide cooling and thus further protects the laser diode.