3.4 System command group
ITC500 / page 68
3.4 System command group
3.4.1 Answer mode
Programming:
Syntax: ":SYST:ANSW FULL"
":SYST:ANSW VALUE"
Default Value: FULL
Description: When switched to "VALUE" the ITC5xx will send only the requested
parameter without designator.
Example: When requesting the actual laser diode current with ":ILD:ACT?"
the ITC5xx will only send [5.123456E-02] instead of
[:ILD:ACT 5.123456E-02]. This is not according to the
IEEE488.2 standard but useful, if you want to increase speed.
Reading:
Syntax: ":SYST:ANSW?"
Response: [:SYST:ANSW FULL]
[:SYST:ANSW VALUE]
Description: Queries the answer mode.
3.4.2 Error-LED mode
Programming:
Syntax: ":SYST:ERRLED OFF" (default)
":SYST:ERRLED ON"
Default Value: OFF
Description: When "ON" the ITC5xx will switch on the interface ERRor LED as
long as there are errors in the error queue. When switched to "OFF"
the interface ERRor LED is dark.
Reading:
Syntax: ":SYST:ERRLED?"
Response: [:SYST:ERRLED OFF]
[:SYST:ERRLED ON]
Description: Queries the Error-LED mode.