3.3 Common commands and queries
ITC500 / page 67
3.3.9 Service-Request-Enable-Register (SRE)
Programming:
Syntax: "*SRE <NR1>"
Valid Range: 0..255
Default Value: 0
Description: Sets the Service-Request-Enable-Register (SRE).
Reading:
Syntax: "*SRE?"
Response: [<NR1>]
Description: Queries the Service-Request-Enable-Register (ESE) and returns the
content in decimal notation. The content is not modified.
3.3.10 Query Status-Byte-Register (STB)
Syntax: "*STB?"
Response: [<NR1>]
Description: Queries the Status-Byte-Register (STB) and returns the content in
decimal notation. Bit 6 (MSS) is set to 0, the other bits keep
unchanged.