3.6 ITC5xx specific commands
ITC500 / page 81
3.6.17 Temperature (only AD590) (TEMP)
Programming:
Syntax: ":TEMP:SET <NR3>"
Valid Range: Depends on the instrument type.
Default Value: 25 °C
Assumption: The temperature sensor is set to AD590.
Description: Sets the TEC temperature. Unit: [°C].
Reading the set temperature:
Syntax: ":TEMP:SET?"
Response: [:TEMP:SET <NR3>]
Assumption: The temperature sensor is set to AD590.
Description: Queries the programmed temperature. Unit: [°C].
Reading the minimum temperature:
Syntax: ":TEMP:MIN?"
Response: [:TEMP:MIN <NR3>]
Assumption: The temperature sensor is set to AD590.
Description: Queries the minimum allowed temperature. Unit: [°C].
Reading the maximum temperature:
Syntax: ":TEMP:MAX?"
Response: [:TEMP:MAX <NR3>]
Assumption: The temperature sensor is set to AD590.
Description: Queries the maximum allowed temperature. Unit: [°C].
Reading the actual temperature:
Syntax: ":TEMP:ACT?"
Response: [:TEMP:ACT <NR3>]
Assumption: The temperature sensor is set to AD590.
Description: Queries the actual temperature. Unit: [°C].