3.7 Error messages of the ITC5xx
ITC500 / page 84
3.7 Error messages of the ITC5xx
Devices following the IEEE488.2 standard provide an error queue storing errors one
by one.
An Error may occur as result of a program message (refer to chapter 3.2.2, “Program
and response messages“ on page 62).
Errors are divided into four categories (refer to chapter 3.8.1, “Standard event status
register (ESR)“ on page 90).
Every query ":SYST:ERR?" will read out one error from the error queue until the
error queue is empty (refer to 3.4.3, “Querying the error queue“ on page 69).
The error queue can keep 32 errors.
If the queue is empty, the error message [0, "No error“] is sent to the PC
3.7.1 General errors
[0, "No errors"]
Category: None
Possible reason: The error queue is empty.
[100, "Unknown command"]
Category: Command Error
Possible reason: ":HELLO WORLD". This string sent to the ITC5xx was not
recognized as valid command.
[101, "Invalid character"]
Category: Command Error
Possible reason: "!". This character sent to the ITC5xx does not belong to the
allowed set of characters.
[102, "Invalid numeric parameter"]
Category: Command Error
Possible reasons: ":ILD:SET 1.2a3". This parameter is not valid.