3.6 ITC5xx specific commands
ITC500 / page 82
3.6.18 Resistance of the thermistor (RESI)
Programming:
Syntax: ":RESI:SET <NR3>"
Valid Range: Depends on the instrument type and the selected temperature sensor.
Default Value: Depends on the selected thermistor range.
Assumption: The temperature sensor is set to thermistor (high or low range).
Description: Sets the TEC thermistor resistance. Unit: [].
Reading the set resistance:
Syntax: ":RESI:SET?"
Response: [:RESI:SET <NR3>]
Assumption: The temperature sensor is set to thermistor (high or low range).
Description: Queries the programmed thermistor resistance. Unit: [].
Reading the minimum resistance:
Syntax: ":RESI:MIN?"
Response: [:RESI:MIN <NR3>]
Assumption: The temperature sensor is set to thermistor (high or low range).
Description: Queries the minimum allowed thermistor resistance. Unit: [].
Reading the maximum resistance:
Syntax: ":RESI:MAX?"
Response: [:RESI:MAX <NR3>]
Assumption: The temperature sensor is set to thermistor (high or low range).
Description: Queries the maximum allowed thermistor resistance. Unit: [].
Reading the actual resistance:
Syntax: ":RESI:ACT?"
Response: [:RESI:ACT <NR3>]
Assumption: The temperature sensor is set to thermistor (high or low range).
Description: Queries the actual thermistor resistance. Unit: [].