3.4 System command group
ITC500 / page 69
3.4.3 Querying the error queue
Syntax: ":SYST:ERR?"
Response: [0, "No Error"]
Description: Queries the error queue of the ITC5xx. The reply consists of the
following sequence: <Error No.>, "<Error text>". If the error queue is
empty: [0, "No error"]is given.
(Please refer to chapter 3.7, “Error messages of the ITC5xx“, starting on page 84)
3.4.4 Averaging rate
Programming:
Syntax: ":SYST:OSR <NR1>"
Valid Range: 0..7
Default Value: 5
Description: Sets the ITC5xx averaging rate (see table below).
Reading:
Syntax: ":SYST:OSR?"
Response: [<NR1>]
Description: Queries the averaging rate.
Sampling rate Resolution Measurement speed
0 11 Bit + sign 18 / s
1 12 Bit + sign 13 / s
2 13 Bit + sign 10 / s
3 14 Bit + sign 8 / s
4 15 Bit + sign 6 / s
5 16 Bit + sign 4 / s
6 17 Bit + sign 3 / s
7 18 Bit + sign 2 / s