1.9 Connecting components
ITC500 / page 26
13
15
14
-
7
5
6
+
9
2
Figure 8 Connecting a TEC
Attention
A TEC element connected with wrong polarity may lead to thermal
runaway and destruction of the connected components.
1.9.6.1 Polarity check of the TEC element
Connect TEC element and temperature sensor. The sensor must have good
thermal contact to the active surface of the TEC element.
Switch on the ITC5xx.
Now follow the steps described in chapter 1.8, “Pre-settings” on page 16
Select the display T
SET
and select a suitable temperature (resistance).
Select the display T
ACT.
Switch on the output (press TEC ON) and observe the actual temperature
(resistance).
If the temperature T
ACT
is continually running away from T
SET
the TEC element is
reverse connected. Change the polarity and repeat the procedure.