2.15 Setup and function of temperature controllers
ITC500 / page 44
To 3. The transient response by setting a new temperature is limited since the heat
transport in the copper bloc is relatively slow. Furthermore, the temperature
slope in the copper must form anew. The sensor (which also possesses a
significant heat capacity) must settle to the laser temperature.
Possibility of optimization: PID adjustment