3.7 Error messages of the ITC5xx
ITC500 / page 86
[303, "Key emulation error"]
Category: Device Error
Possible reason: Internal communication problem. Please contact Thorlabs.
[400, "Too many errors"]
Category: Device Error
Possible reason: Error queue overflow (32 errors max).
[410, "Query interrupted"]
Category: Query Error
Possible reason: More than one query sent to the ITC5xx before the read
command.
[420, "Query unterminated"]
Category: Query Error
Possible reason: There is no data in the output buffer.
[500, "IEEE488 receive buffer overflow"]
Category: Device Error
Possible reason: The string sent to the ITC5xx was too long for the IEEE488
receive buffer (250 char max).