EasyManuals Logo

THORLABS ITC510 User Manual

THORLABS ITC510
123 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #94 background imageLoading...
Page #94 background image
3.7 Error messages of the ITC5xx
ITC500 / page 86
[303, "Key emulation error"]
Category: Device Error
Possible reason: Internal communication problem. Please contact Thorlabs.
[400, "Too many errors"]
Category: Device Error
Possible reason: Error queue overflow (32 errors max).
[410, "Query interrupted"]
Category: Query Error
Possible reason: More than one query sent to the ITC5xx before the read
command.
[420, "Query unterminated"]
Category: Query Error
Possible reason: There is no data in the output buffer.
[500, "IEEE488 receive buffer overflow"]
Category: Device Error
Possible reason: The string sent to the ITC5xx was too long for the IEEE488
receive buffer (250 char max).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the THORLABS ITC510 and is the answer not in the manual?

THORLABS ITC510 Specifications

General IconGeneral
BrandTHORLABS
ModelITC510
CategoryController
LanguageEnglish

Related product manuals