FFS2000 Series Chapter 2: Safety
Page 2 TTN047338-D02
Chapter 2 Safety
All statements regarding safety of operation and technical data in this instruction manual will only apply when the
unit is operated correctly.
SHOCK WARNING
Unplug the power cord before servicing the unit. Do not operate the unit without all covers and items
properly installed.
CAUTION
Before connecting the AC power cord, make sure the source voltage is between 85 and 265 VAC, 47-
63 Hz. Input voltages outside these ranges may result in damage to the unit.
Ensure that the Personal Computer voltage switch is set appropriately for the local supply.
CAUTION
The colorless solution used to clean the fibers may either be acetone (> 99%) or isopropanol (>
99.5%). Both are volatile and flammable. Ensure that the work area is well ventilated to avoid
excessive inhalation of the solution being used. Do not ingest and avoid contact with the skin and
eyes. Prolonged skin contact may result in severe irritation.
CAUTION
The recoat material is a flammable, toxic, acrylate compound. Avoid contact with the skin and eyes.
WARNING
Argon is a chemically inert gas. It is colorless, odorless, tasteless, non-flammable, non-corrosive and
non-toxic. However, the work area should be well ventilated so as to ensure that the correct oxygen
level is maintained.
WARNING
Ensure that the operator is properly trained for the handling of compressed gases and regulators.
Small transparent fiber shards may be present in and around the work area, and as such, the
necessary measures should be taken to ensure the safety of the employee(s).
!
!
!
!
!
!
!
!
!
!