78
NOTE
A moisture offset simply adds or subtracts a set value
to the moisture probe reading using the enabled
calibration (usually a factory calibration). The
Moisture Probe Calibration (page 515) recalibrates
the probe for that particular soil allowing a material
specific profile to be created.
Then enter the Wet Density value from the initial measurement
site. Press Enter to accept.
Last enter the True Moisture percent (%M) resulting from the
lab moisture analysis (oven dry or Speedy for example) of the
soil taken from the initial site.
The gauge then displays the new moisture offset value and
returns to the Offset menu.
Input Wet Density
WD= 0.0 pcf
<ENTER> to Accept
Input Probe
Moisture
M= 0.00 pcf
<ENTER> to Accept
Input True Moisture
%M= 0.00 %Wt
<ENTER> to Accept