EasyManuals Logo

Troxler EGauge Combo 4540 User Manual

Troxler EGauge Combo 4540
188 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #164 background imageLoading...
Page #164 background image
D-8
Emergency Procedures
If the nuclear gauge is lost or stolen, then immediately notify the
gauge owner. The gauge owner should complete the emergency
contact information on the lines furnished below. This
information should be readily available to the gauge operator at
all times.
If a gauge is damaged, then follow the steps below:
1. Locate the gauge and/or sources.
2. Do not touch or move the gauge.
3. Immediately rope off an area around the nuclear gauge
and/or sources. A radius of 6 ft. (2 m) is sufficient. Do not
leave the area unattended.
4. Keep all unauthorized personnel from the nuclear gauge.
5. The gauge user should perform a visual inspection of the
nuclear gauge to determine if the source housing or
shielding has been damaged.
6. Contact the gauge owner (name and number given at the
beginning of this section). Provide the gauge owner with the
following:
The date, time, and location of the accident
The gauge model and serial number
The nature of the accident
The location and condition of the gauge and/or source
Gauge Owners Name
Gauge Owners Phone
Additional Contact
Additional Contact

Table of Contents

Other manuals for Troxler EGauge Combo 4540

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Troxler EGauge Combo 4540 and is the answer not in the manual?

Troxler EGauge Combo 4540 Specifications

General IconGeneral
BrandTroxler
ModelEGauge Combo 4540
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals