74
Density Offset
To access the Density Offset functions, press 1 at the Offset
menu.
The gauge displays the current density offset on the second line.
To enable the displayed density offset, press 1. The
gauge enables the offset, displays a brief confirmation
message, and returns to the Ready screen.
To disable the density offset, press 2. The gauge
disables the offset, displays a brief confirmation message,
and returns to the Ready screen.
NOTE
The density offset is also disabled when the gauge is
turned off for more than 10 seconds
To enter a new density offset, press 3. The gauge
prompts for the density offset as shown:
The density offset value is determined by calculating the
difference between the gauge wet density result and an
alternative wet density measurement (Ex. Sand cone or
drive cylinder) performed on the particular soil being
measured.
DENSITY OFFSET
#.# pcf
1. Enable 2. Disable
3. Change Offset
DENSITY OFFSET
#.# pcf
Select (+/-)
Input and <ENTER>

Table of Contents

Other manuals for Troxler EGauge Combo 4540

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Troxler EGauge Combo 4540 and is the answer not in the manual?

Troxler EGauge Combo 4540 Specifications

General IconGeneral
BrandTroxler
ModelEGauge Combo 4540
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals