92
5 COMMUNICATIONS
92
COMMUNICATIONS
Chapter
5
Read the hold results.
HD1L
Delimiter
SP - 3
Space (20H)
Hyphen (2dH)
First Section No. to read (1 to 5)
1 to 5: Section
Host
(Transmission to the F381A)
Last Section No. to read (1 to 5)
1 to 5: Section
Specify so that the last the first.
If the first section = the last section,
only the specified section can be read.
STX
1HDLSP -
3
F381A’s return
(In normal processing)
Command received by the F381A
.0100-
20210.
,,
2-00
Load result
-: Section not in use
1: OK
2: HI
3: LO
4: H/L
5: NG
Displacement or Time*
If there is no hold point in the specified section, hyphens (2dH) enter into Load
and Displacement (Time), and if “Always” is set in the specified section,
hyphens (2dH) also enter into Load and Displacement (Time).
Displacement result
Data of the specified first section
.0202-
20150.
,,
2-00
2nd data
----3-
------
,,
1---
3rd data
Delimiter
-: Section not in use
or X-axis; time
1: OK
2: HI
3: LO
4: H/L
5: NG
STX
Delimiter
N1G
F381A’s return
(In abnormal condition)
Header (If there is no header, no data will result.)
NG No.
1: Receipt of a statement different from the command format
2: Attempt to write a set value outside the setting range
Load
-9999 to +9999
(with a decimal point)
* Displacement (Front): 0 to 10235 (with a decimal point)
Displacement (Back): -10235 to 0 (with a decimal point)
Time: +00000 to +51175 (with a decimal point)
The unit of time differs according to Measurement Length, X-axis Setting.
0.5s: Unit ms
1.0s: Unit ms
2.0s: Unit ms
5.0s: Unit ms
10.00s: Unit s