EasyManua.ls Logo

Unipulse F381A - Connection of a Lightning Surge Protector

Unipulse F381A
164 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
153
9 SUPPLEMENTS
153
SUPPLEMENTS
Chapter
9
Connection of a lightning surge protector
Attach a lightning surge protector against lightning surge.Also, for using the Ethernet interface,
attach a lightning surge protector for LAN separately.
EMC Directive EN61000-4-5 (Lightning Surge Immunity) is met by the F381A body in
combination with a lightning surge protector and a lightning surge protector for LAN .
● At the Ethernet interface use
Lightning surge protector
* PT-BE/FM and PT 2-PE/S-24AC-ST are registered
trademarks of PHOENIX CONTACT.
PT-BE/FM
PT 2-PE/S-24AC-ST
(DIN rail mount type)
F381A body
D-LAN-CAT.5E is registered trademarks of PHOENIX
CONTACT.
F381A body
Lightning surge protector for LAN
 D-LAN-CAT.5E
Ethernet
Connection apparatus

Table of Contents

Related product manuals