142
9 SUPPLEMENTS
142
SUPPLEMENTS
Chapter
9
9-2. Setting item list
■First Setting
*1: Function-specific set values (calibration values, etc.); currently set values remain even if initialization is
executed.
*2: [NOV RAM : N] [SRAM : S]
Item Set value Input range (display range)
Initial
value
Memory
*2
Remarks
Y-axis Setting
Exc. Voltage
*1
0 : 2.5 1 : 10 [V] 2.5V N
Unit
*1
See "9-3 Unit setting list" on p.147 kN N
Zero Cal.
*1
-3.333 to 3.333 [mV/V] 0.000mV/V N
Equiv. Cal.
*1
(rated output)
-9.999 to 9.999 [mV/V] 3.000mV/V
N
Equiv. Cal.
*1
(rated capacity)
-9999 to 9999 50.00kN
N
Actual Cal.
*1
(Load value)
-9999 to 9999 50.00kN
N
Overload 0 to 9999 99.99kN N
Increment
*1
0 : 1 1 : 2
2 : 5 3 : 10
0.01kN
N
Analog Filter 0 : 10 1 : 30
2 : 100 3 : 300 [Hz]
300Hz
N
Digital Filter 0, 2 to 999 [Times] 0 Times N
DZ Limit 0 to 9999 99.99kN N
X-axis Setting
(standard)
Time/DPM Change 0: Time 1: Displacement Time N
Phase Select A/B-Phase 1: Only A-Phase A/B-Phase N
Wave Reference 0: Front 1: Back Front N
Unit
*1
See "9-3 Unit setting list" on p.147 mm N
Zero Cal.
*1
0 to 1000000 0 N
Equiv. Cal.
*1
(number of pulses)
1 to 1000000 1000
N
Equiv. Cal
*1
(displacement value)
-99999 to 99999 10.00mm
N
Actual Cal.
*1
(displacement value)
-9999 to 32000 10.00mm
N
Measure Length Time/Displacement Change: Time
0 : 0.5 1 : 1.0
2 : 2.0 3 : 5.0
4 : 10.0 [s]
Time/Displacement Change: Displacement
0 : 2000 1 : 4000
2 : 6000 3 : 8000
4 : 10000
2.0s
60.00mm
N
DPM Positioning -9999 to 32000 0.00mm N
Digital Filter 0, 2 to 999 [Times] 0 Times N