14
2 INSTALLATION AND CONNECTION
14
INSTALLATION AND CONNECTION
Chapter
2
■Displacement sensor connection (when the voltage input option is mounted)
Up to ±5V voltage output type sensor of the contact type, eddy current type, laser type, etc., can be
connected. (The voltage input option is order-specified.)
■Displacement sensor connection (when the line driver input option is mounted)
It can be connected with Line driver output-type sensor (Incremental type) in accordance with RS-
422 like the contact-type, eddy current-type or laser-type etc.
(The Line driver input option is order-specified.)
8
9
10
11
12
VIN
GND
Voltage output
Output within ±5V
+
-
F381A body
F381A
Outside
Inside
type sensor
F.G.
VIN
GND
* Please wire for the power supply of the sensor separately.
A+
B+
Line driver
F381A body
F381A
Outside
Inside
output-type sensor
F.G.
8
A-
B-
9
10
11
12
A+(PCA、φA)
B+(PCB、φB)
A-(A/、A、*PCA、φA)
B-(B/、B、*PCB、φB)
* Please wire for the power supply of the sensor separately.