88
5 COMMUNICATIONS
88
COMMUNICATIONS
Chapter
5
Reading Y-axis Start Point after Work No. is written normally
- Set value write command (First Setting)
Example: Writing 1:10 [V] to Excitation Voltage, Y-axis Setting, First Setting
1
STX
Delimiter
NG
R201
Delimiter
0
Host
F381A’s return
(In normal processing)
(Transmission to the F381A)
Item No. (See the list of set value commands.)
Y-axis Start Point in the example
Item group (See the list of set value commands.)
Display Range Setting in the example
STX
Delimiter
10R2 010+00 000
F381A’s return
(In abnormal condition)
Header (If there is no header, no data will result.)
NG No.
1: Receipt of a statement different from the command format
Data of the specified set value
Command received by the F381A
Header (If there is no header, no data will result.)
Host
F381A’s return
(In normal processing)
Command received by the F381A
Header (If there is no header, no data will result.)
(Transmission to the F381A)
Item No. (See the list of set value commands.)
Excitation Voltage in the example
Item group (See the list of set value commands.)
Y-axis Setting in the example
STX
Delimiter
00W1 000+00 010
00+0000001W0 1
Delimiter
Data to write
“1” in the example
STX
Delimiter
N1G
F381A’s return
(In abnormal condition)
Header (If there is no header, no data will result.)
NG No.
1: Receipt of a statement different from the command format
2: Attempt to write a set value outside the setting range
4: Attempt to write a set value through communications
during measurement
5: Attempt to write while protecting