47
4 SETTINGS AND CALIBRATIONS
47
SETTINGS AND CALIBRATIONS
Chapter
4
4-5. Measurement start condition setting
Set measurement conditions by Measurement Start Condition Setting. The Measurement Start
Condition is as follows:
Main screen→Setting→Work Setting→Start Setting
1. Measurement Start Condition
Set the measurement start timing.
Main screen→Setting→Work Setting→Start Setting→Start Condition
①Only External Input
Measurement starts when the external input signal
“Start” turns OFF →ON →OFF.
Setting of “2. Measurement Start Level” is disabled.
②External Input + Load
Measurement starts when the load crosses the
Measurement Start Level after the external input signal
“Start” turns OFF → ON → OFF.
Load
Measurement
Time or displacement
Start
OFF
ON
Section
Load
Time or displacement
Start
OFF
ON
Start
load
Measurement
Section
Start level
Load
Start point Start point
×
Not crossing
×
Not crossing
〇
Crossing in the
negative-to-
〇
Crossing in the
positive-to-
positive direction negative direction