48
4 SETTINGS AND CALIBRATIONS
48
SETTINGS AND CALIBRATIONS
Chapter
4
External Input + Displacement
Measurement starts when the displacement crosses the
Measurement Start Level after the external input signal
“Start” turns OFF ON OFF.
2. Measurement Start Level
Set load or displacement when the Measurement Start Condition
is “External Input + Load” or “External Input + Displacement.”
Setting range
External Input + Load (-9999 to +9999)
Operation
External Input + Displacement (-9999 to +32000)
Main screenSettingWork SettingStart SettingStart Level
3. Measurement Stop Condition
Set the Measurement Stop Condition.
Operation
Main screenSettingWork SettingStart SettingStop Condition
Load
Displace-
Start
OFF
ON
Start
displacement
Measurement
Section
ment
Start
level
Start point Start point
×
Not crossing
×
Not crossing
Crossing in the
negative-to-
Crossing in the
positive-to-
Displace-
Start
level
Start
level
Start
level
positive direction negative direction
ment
Displace-
ment
Displace-
ment
Displace-
ment
The operation of turning the external input signal OFF ON OFF is the
same as pressing on the Main screen.
Point