- Zero calibration command
Example: To perform zero calibration (load)
- Actual load calibration command
Example: To perform actual load calibration (load)
C100Delimiter0
Host
(Transmission to the F381A)
Item No. (See the list of Zero Calibration commands.)
Zero Calibration in the example
Item group (See the list of Zero Calibration commands.)
Y-axis Setting in the example
F381A’s return
(In normal processing)
Command received by the F381A
Header (If there is no header, no data will result.)
STX Delimiter0C100
STX
Delimiter
N1G
F381A’s return
(In abnormal condition)
Header (If there is no header, no data will result.)
NG No.
1: Receipt of a statement different from the command format
2: Attempt to write a set value outside the setting range
3: Attempt to perform calibration with actual load outside the calibration range
4: Attempt to write a set value through communications during measurement
5: Attempt to write while protecting
Host
(Transmission to the F381A)
Item No. (See the list of Actual Calibration commands.)
Actual Calibration in the example
Item group (See the list of Actual Calibration commands.)
Y-axis Setting in the example
00+1000100C01
Delimiter
Load-applied value
“1000” in the example
F381A’s return
(In normal processing)
Command received by the F381A
Header (If there is no header, no data will result.)
STX
Delimiter
11C1 000+00 000
STX
Delimiter
N1G
F381A’s return
(In abnormal condition)
Header (If there is no header, no data will result.)
NG No.
1: Receipt of a statement different from the command format
2: Attempt to write a set value outside the setting range
3: Attempt to perform calibration with actual load outside the calibration range
4: Attempt to write a set value through communications during measurement
5: Attempt to write while protecting