72
4 SETTINGS AND CALIBRATIONS
72
SETTINGS AND CALIBRATIONS
Chapter
4
4-11.Measuring work specification
Up to 16 types of work set values can be stored in the F381A, and the work No. (measuring work)
used for measurement must be specified.
For this, the external input signals “Work 1 to Work 8” are used.
Normally, if there is no input to “Work 1 to Work 8,” the Work 0 set value is selected, while if
“Work 1 to Work 8” are in the following conditions, each work is selected.
(0:Open, 1:ON)
Work 8 Work 4 Work 2 Work 1
Main work display
(measuring work)
0000 0
0001 1
0010 2
0011 3
0100 4
0101 5
0110 6
0111 7
1000 8
1001 9
1010 10
1011 11
1100 12
1101 13
1110 14
1111 15
Control connector
A9, A10
B9, B10
COM (Input)
B1 Work 1
B2 Work 2
B3 Work 4
B4 Work 8
Measuring work is accepted when the Start
input signal is turned ON.
Even if the measuring work is changed
during measurement, it is ignored.
Point