29
3 SCREENS AND OPERATIONS
29
SCREENS AND OPERATIONS
Chapter
3
Result display : The result is displayed.
The waveform comparison and the hold results of each section are displayed
overall according to the priority.
* The displacement result is not displayed if X axis is time.
Indicated value display:
Displays the indicated value selected by (digital value of the load sensor/
displacement sensor input), and also displays load sensor/displacement sensor errors.
* When the indicated value selected by is Section 1 - 5, if there is no hold
value, “-----” is displayed.
Work display: Displays the present work No. specified by external input “Work.”
Selected indicated value display: Displays the currently-selected indicated value.
Present........ Present value
1 ................... Section 1
2 ................... Section 2
3 ................... Section 3
4 ................... Section 4
5 ................... Section 5
Keys
: Moves to a result screen.
There are two types of screens: [Result (Single)] and [Result (List)].
In moving from the Main screen, the previously displayed screen appears.
(The default display is [Result (Single)].)
: Moves to the Wave Check screen.
: Moves to the Setting screen.
: Simply makes the same movement as the Reset signal.
Also, when the error display appears, this key clears it after release.
: Simply makes the same movement as the Start signal.
This key is valid when the status display is Complete or Wait St.
When pressed, it changes to .
Priority Result Judgment condition
1H/LHI/LO limit over (HI limit over and LO limit over)
2
HI HI limit over (judgment point > HI limit set value)
LO LO limit over (judgment point < LO limit set value)
3NG
<Hold operation>
- Measurement did not reach the section.
- Failure to detect the judgment point.
* If the setting is such that Start Point of Section n = End Point of
Section n = Start Point of Section n+1, the result of the present
section is always “NG.”
<Waveform comparison operation>
- Measurement did not reach the Compare Area.
- Measurement did not reach the Reference Point of Relative
Comparison when Relative Comparison is made.
4OK
Normal
(LO limit set value
Judgment point
HI limit set value)