EasyManua.ls Logo

Unipulse F381A - Page 54

Unipulse F381A
164 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
45
4 SETTINGS AND CALIBRATIONS
45
SETTINGS AND CALIBRATIONS
Chapter
4
Equivalent Input Calibration
(when the voltage input option is used)
Input the relationship between the voltage value (rated output)
and displacement value (display value) determined from the data
sheet of the voltage sensor. Also, set the decimal place here.
Input the decimal point together when inputting the display
value.
Setting range
Rated output (voltage value) (-9.999 to 9.999 V)
Operation
Display value (displacement value) (-99999 to 99999)
Main screenSetting First SettingX-axis SettingNextEquiv. Cal.
7. Actual Load Calibration
Move the cylinder, and input its position. Also, set the decimal
place here. Input the decimal point together when inputting the
calibration value.
Setting range
Calibration value (-9999 to 32000)
Operation
Display value (display only)
Main screenSetting First SettingX-axis SettingNextActual Cal.
8. Measurement Length
Set the length for waveform acquisition.
Operation
Setting range (2000, 4000, 6000, 8000, 10000)
Main screenSetting First SettingX-axis SettingNextMeasure Length
9. Displacement Positioning
Set the position for displacement adjustment at power-on, on the
Main screen, and by external input.
Operation
Setting range (-9999 to 32000)
Main screen
Setting
First Setting
X-axis Setting
Next
Next
DPM Positioning
10.Analog Filter (when the voltage input option is used)
The low-pass filter of the set cut-off frequency is inserted with
respect to the input of the voltage sensor.
Operation
Setting range (10, 30, 100, 300 Hz)
Main screen
Setting
First Setting
X-axis Setting
Next
Next
Analog Filter

Table of Contents

Related product manuals