67
4 SETTINGS AND CALIBRATIONS
67
SETTINGS AND CALIBRATIONS
Chapter
4
1. Registration of Reference Waveform
Input and register reference waveform(s).
Operation
Main screen Setting Work Setting Wave Setting Reference Wave
Not to use the currently-registered waveform(s), press
, and select .
Satisfy the Measurement Start Condition, and start measurement.
Satisfy the Measurement Stop Condition, and stop measurement.
A measurement waveform is displayed.
Now, to delete the measurement waveform, press , and
select .
To input a waveform again, repeat to .
Pressing registers the reference waveform(s).
In the case of one reference waveform, it is
recognized that the HI limit waveform and LO limit
waveform are the same.
If two or more waveforms are input, the upper limit
of all waveforms becomes the HI limit waveform,
and the lower limit of all waveforms becomes the
LO limit waveform.
Load
1st waveform
Time or displacement
Load
1st waveform
Time or displacement
2nd waveform
3rd waveform